Adverse outcomes of interrupted precordial compression during automated defibrillation.
Circulation 2002 July 17
BACKGROUND: Current versions of automated external defibrillators (AEDs) require frequent stopping of chest compression for rhythm analyses and capacity charging. The present study was undertaken to evaluate the effects of these interruptions during the operation of AEDs.
METHODS AND RESULTS: Ventricular fibrillation was electrically induced in 20 male domestic swine weighing between 37.5 and 43 kg that were untreated for 7 minutes before CPR was started. Defibrillation was attempted with up to 3 sequential 150-J biphasic shocks, but each was preceded by 3-, 10-, 15-, or 20-second interruptions of chest compression. The interruptions corresponded to those that were mandated by commercially marketed AEDs for rhythm analyses and capacitor charge. The sequence of up to 3 electrical shocks and delays were repeated at 1-minute intervals until the animals were successfully resuscitated or for a total of 15 minutes. Spontaneous circulation was restored in each of 5 animals in which precordial compression was delayed for 3 seconds before the delivery of the first and subsequent shocks but in none of the animals in which the delay was >15 seconds before the delivery of the first and subsequent shocks. Longer intervals of CPR interventions were required, and there was correspondingly greater failure of resuscitation in close relationship to increasing delays. The durations of interruptions were inversely related to the durations of subthreshold levels of coronary perfusion pressure. Postresuscitation arterial pressure and left ventricular ejection fraction were more severely impaired with increasing delays.
CONCLUSIONS: Interruptions of precordial compression for rhythm analyses that exceed 15 seconds before each shock compromise the outcome of CPR and increase the severity of postresuscitation myocardial dysfunction.
METHODS AND RESULTS: Ventricular fibrillation was electrically induced in 20 male domestic swine weighing between 37.5 and 43 kg that were untreated for 7 minutes before CPR was started. Defibrillation was attempted with up to 3 sequential 150-J biphasic shocks, but each was preceded by 3-, 10-, 15-, or 20-second interruptions of chest compression. The interruptions corresponded to those that were mandated by commercially marketed AEDs for rhythm analyses and capacitor charge. The sequence of up to 3 electrical shocks and delays were repeated at 1-minute intervals until the animals were successfully resuscitated or for a total of 15 minutes. Spontaneous circulation was restored in each of 5 animals in which precordial compression was delayed for 3 seconds before the delivery of the first and subsequent shocks but in none of the animals in which the delay was >15 seconds before the delivery of the first and subsequent shocks. Longer intervals of CPR interventions were required, and there was correspondingly greater failure of resuscitation in close relationship to increasing delays. The durations of interruptions were inversely related to the durations of subthreshold levels of coronary perfusion pressure. Postresuscitation arterial pressure and left ventricular ejection fraction were more severely impaired with increasing delays.
CONCLUSIONS: Interruptions of precordial compression for rhythm analyses that exceed 15 seconds before each shock compromise the outcome of CPR and increase the severity of postresuscitation myocardial dysfunction.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app